YourLawyer.com 1-800-LAW-INFO (1-800-529-4636)

St Jude Defibrillators


AddThis Social Bookmark Button

St. Jude Riata Defibrillator Lead Wires Perforation Problems

In November 2007, reports emerged that St. Jude Riata Defibrillator Lead wires had perforated the hearts of some patients.   That month, the medical journal “Pace” published a report detailing four instances where the St. Jude Riata Defibrillator Lead wire detached and perforated the heart wall.  In one instance, the defective St. Jude Riata Defibrillator Lead wire not only perforated the heart, but it nearly poked through a patient’s skin. 

When a defibrillator lead wire becomes detached from the heart, the device will fail to emit needed electrical shocks to the heart, leaving a patient at risk of death.   If the detached St. Jude Riata Defibrillator Lead perforates the heart tissue, the heart can bleed into the pericardial sac around it.  This causes a lethal condition called cardiac tamponade in which pressure builds around the heart, preventing it from beating effectively.  In an editorial accompanying the “Pace” article, Dr. Stephen Vlay, a cardiologist for Stony Brook University, wrote that the problems with the St. Jude Riata Defibrillator Leads could be due to an “inherent design flaw, at least in some models of the Riata lead.” 

Earlier in 2007, the medical journal “Heart Rhythm” reported that the St. Jude Riata Defibrillator Lead had a higher perforation rate than what had been stated by the manufacturer.  According to that article, doctors at Massachusetts General Hospital reported a perforation rate of 3.8% - or 5 out of 130 – for the St. Jude Riata Defibrillator Lead.  Doctors from New York Hospital Queens also reported to “Heart Rhythm” that of 59 St. Jude Riata Defibrillator Leads implanted there, five had perforated.   New York Hospital Queens has since stopped using the defective St. Jude Riata Defibrillator Lead.

St. Jude Defibrillators

St. Jude Medical Inc., the maker of implantable devices that regulate heart rhythms, said it has identified a memory chip problem in a small number of some of its older devices. While no deaths or serious injuries have been reported to the company as a result of the problem, St. Jude said in a government filing that the U.S. Food and Drug Administration may classify the problem as a recall.

The problem which stems from a memory chip that St. Jude used through 2002 that is susceptible to background radiation has not been seen in any of the devices the company currently sells. The St. Jude implantable cardioverter defibrillators (ICDs) that may be affected by the radiation problem are certain older generations of the company's Photon DR, Photon Micro VR/DR and Atlas VR/DR models.

So far, only 60 out of an approximate 36,000 devices have been found to be affected, the company said in the filing. Nearly 26,000 of the devices remain in patients.

St. Jude Medical Inc. has notified doctors and federal regulators of a software problem in some models of its implantable defibrillators that could cause the heart-shocking device to malfunction. About 39,000 patients are affected by this news.

ICDs are implanted devices the size of a stopwatch that are placed in the upper chest and shock or pace an irregularer heart beat back into rhythm. St. Jude said it discovered the two "anomalies" during a routine product evaluation.

In ICDs past their mid-life, a series of shocks might be skipped as the device delivers its routine of multiple shocks to revitalize the patients heart. The first shock would be delivered, but the device might deliver fewer than the maximum of six shocks per episode. The devices on average last four to seven years before replacement. A second problem could cause a temporary increase in the device's pacing rate. In addition to its shocking therapy, some patients need an ICD that can pace their heart, as well.

The affected models include: Epic DR/HF (V-233/V-337/V-338), Epic Plus DR/VR/HF(V-236/V-239/V-196/V-239T/V-239T/V-196T/V-350), Atlas DR (V-242), and Atlas Plus DR/VR/HF (V-243/V-193/V-193C/V-340/V-341/V-343).

If you or a loved one suffered injuries from a St Jude defibrillator, please fill out the form at the right for a free case evaluation by a qualified defective medical device attorney.
St Jude DefibrillatorsRSS Feed

Heart Device Patients Often Unaware of Recalls

May 16, 2008 | Parker Waichman Alonso LLP
It seems that many patients who have been implanted with heart devices are not aware of recent recalls and do not understand the dangers they might face, this according to two new surveys.  Although the surveys are small, they do indicate a lack of understanding among doctors and patients about heart disease and the devices used to treat it, said Dr. Bruce Wilkoff, director of cardiac and tachyarrhythmia devices at Cleveland Clinic Wilkoff, who is familiar with the surveys. "There are...

Medtronic Sprint Fidelis, St. Jude Riata Defibrillator Lead Wires Endanger Patients

Nov 19, 2007 | Parker Waichman Alonso LLP
People with a Medtronic or St. Jude implantable defibrillators should know that  some of these defibrillators’ lead wires—specifically in the Medtronic Sprint Fidelis and St. Jude Riata lead wires—have fractured, perforated patient heart walls, or caused device failure.  In some cases, these malfunctions of the Medtronic Sprint Fidelis Lead and the St. Jude Riata Defibrillator Lead  have led to additional cardiac procedures or complications and even...

St. Jude Defends Riata Defibrillator Lead, But Expert Skeptical

Nov 13, 2007 | Parker Waichman Alonso LLP
St. Jude Medical Inc. is trying to undo the damage caused by reports that its Riata Defibrillator Lead wires have perforated the hearts of some patients.  The company claims that reports that the St. Jude Riata Defibrillator Lead wire is more prone to cause heart perforations are overstated, and that its own data shows the Riata Lead is safe.  However, at least one heart expert has disputed St. Jude’s contention, telling the Wall Street Journal that the company’s data...

St. Jude Riata Defibrillator Lead Wires Perforating Patients' Hearts

Nov 12, 2007 | Parker Waichman Alonso LLP
St. Jude Riata Defibrillator Lead wires have been detaching from patients’ hearts and perforating the heart wall, according to reports published in a major medical journal.   While such malfunctions can happen with any defibrillator lead, evidence is emerging that such detachment and perforation is occurring more frequently than expected with the St. Jude Riata Defibrillator Leads.A defibrillator lead is a thin wire that attaches an implantable defibrillator to the heart. ...

Health Canada Endorsed Important Safety information on certain St. Jude Medical implantable cardiac defibrillators

Nov 6, 2005 | Health Canada
Dear Doctor:This letter is written to provide you with important safety information regarding a memory chip component supplied to us and used in a limited number of St. Jude Medical ICD products within the Photon and Photon Micro device families and in certain Atlas devices.Specifically, we have identified that a particular vendor-supplied memory chip can be affected at a low frequency rate by background levels of atmospheric ionizing cosmic radiation ("background cosmic radiation")....

More St Jude Defibrillators News

St. Jude Defibrillators
* Denotes required field.

Title

* First Name

* Last Name

* Email Address

* Phone Number

Cell Phone Number

Office Phone Number

Street Address

Apartment/Suite

City

State

Zip Code

Please provide the best method and times to contact you:

Date of birth of person injured
(mm-dd-yyyy):

Date the device was implanted:

Is the device still implanted?


What medical condition prompted the use of the device?

Please describe any problems or injuries caused by the device:

Other Info:

No Yes, I agree to the Parker Waichman Alonso LLP disclaimers.Click here to review all.

Yes, I would like to receive the Parker Waichman Alonso LLP monthly newsletter, InjuryAlert.

please do not fill out the field below.

Home | Defective Drugs | Medical Devices | Toxic Substances | Accidents | Product Liability | Malpractice | Diseases
Nursing Home Negligence | Food Poisoning | Other Topics | Contact
Statement of Clients' Rights | Site Map | ReNu with MoistureLoc | Vioxx | Mesothelioma | Permax | Dostinex | Composix Kugel Mesh X Large Patch
Ortho Evra | Fosamax | Personal Injury Lawyer | Fusarium Keratitis | Stevens Johnson Syndrome

© 2002-2008 YourLawyer.com. All Rights Reserved.

Please note that you are not considered a client until you have signed a retainer agreement and your case has been accepted by us.
Prior results do not guarantee or predict a similar outcome with respect to any future matter.
Attorney Advertising

Parker Waichman Personal Injury News